Recall of BUSH SL URETERAL ILLUMINATING CATHETER SET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by COOK (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74503
  • Event Risk Class
    II
  • Event Initiated Date
    2017-03-07
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The proximal black component that connects to the distal transparent component of the catheter may overheat and melt. there have been no reports of overheating along the transparent component of the catheter which is the part of the device that comes into contact with the patient's urethra bladder and ureter. potential adverse events that may occur if these devices overheat and come into direct contact with skin include burns to the skin.

Device

  • Model / Serial
    Model Catalog: J-BICS-058020 (Lot serial: > 10 pls contact mfr.); Model Catalog: 084120 (Lot serial: > 10 pls contact mfr.); Model Catalog: 084510 (Lot serial: > 10 pls contact mfr.)
  • Product Description
    Cook Bush Ureteral Illuminating Catheter Set
  • Manufacturer

Manufacturer