Recall of BECKMAN COULTER PK7300 AUTOMATED MICROPLATE SYSTEM (DONOR SCREENING) - INSTRUMENT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    80095
  • Event Risk Class
    III
  • Event Initiated Date
    2012-08-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The reagent transfer (ed assembly) probe aspiration height was adjusted during a service visit to a value below the acceptable range for the aspiration height (116.45 - 116.95 mm) on one pk7300 device in canada.

Device

  • Model / Serial
    Model Catalog: PK7300 (Lot serial: 9110205)
  • Product Description
    PK7300 AUTOMATED MICROPLATE SYSTEM
  • Manufacturer

Manufacturer