Recall of BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - PREALBUMIN ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16707
  • Event Risk Class
    III
  • Event Initiated Date
    2012-02-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Currently the prealbumin reagent osr6175 does not meet the ifu lipemic (intralipid) interference claim of "no significant interference up to 500 mg/dl intralipid". in-house studies have demonstrated that the effect of intralipid on the prealbumin reagent varies and has been observed to cause a maximum negative interference of -27% up to 500 mg/dl intralipid on prealbumin recovery.