Recall of BD VACUTAINER PLUS PLASTIC PLASMA SEPARATOR TUBES (PST)

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECTON DICKINSON CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    73859
  • Event Risk Class
    II
  • Event Initiated Date
    2013-04-29
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A product defect named "lip-out" was detected for bd vacutainer pst tubes. it consists of gaps of various sizes in the structure of the tube wall that is present at the lip of the tube. these gaps are covered by the stopper and hemogard closure so the tube can be evacuated and used for blood collection. customer complaints cite jamming of preanalytical automation (tube decappers) and leakage of blood in the specimen transport bag.

Device

  • Model / Serial
    Model Catalog: 367962 (Lot serial: >10 contact manufacturer); Model Catalog: 367960 (Lot serial: >10 contact manufacturer); Model Catalog: 367961 (Lot serial: >10 contact manufacturer)
  • Product Description
    BD Vacutainer Plus Plastic Plasma Separator Tube (PST)
  • Manufacturer

Manufacturer