Recall of BARIATRIC BED

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by INVACARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31103
  • Event Risk Class
    II
  • Event Initiated Date
    2012-09-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    There is the potential for misconnecting or miskeying the electrical connector (between the head motor extension cable and the foot section) during initial set-up or installation of the product. this results in immediate and permanent failure on first use and has the potential to cause a fire.

Device

  • Model / Serial
    Model Catalog: BARPKGIVC-1633 (Lot serial: 01Fxxxxxx to 11KF024680); Model Catalog: BARPKG (Lot serial: 01Fxxxxxx to 11KF024680); Model Catalog: BARPKGCA (Lot serial: 01Fxxxxxx to 11KF024680); Model Catalog: BAR600IVC (Lot serial: 01Fxxxxxx to 11KF024680); Model Catalog: BAR600 (Lot serial: 01Fxxxxxx to 11KF024680)
  • Product Description
    BARIATRIC BED
  • Manufacturer

Manufacturer