Recall of AXIOM VERTIX MD TRAUMA

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    129722
  • Event Risk Class
    II
  • Event Initiated Date
    2014-03-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    In rare cases steel cables inside the lifting column of the system can fail without triggering the safety lock. this can result in the system u-arm dropping down unexpectedly during movement in the vertical direction. since the u-arm weight is approximately 100 kg failure of the steel cables may result in serious injury to patients.

Device

  • Model / Serial
    Model Catalog: 8908290 (Lot serial: SN 1041 1058)
  • Product Description
    AXIOM VERTIX MD TRAUMA
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC