Recall of AXIOM ARTIS BA

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18818
  • Event Risk Class
    II
  • Event Initiated Date
    2016-07-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens is initiating a field corrective action as during routine surveillance activities it was identified that contaminants in the form of biomass can develop in the cooling system of artis systems which can result in damage to the pump system. biomass is developing in the ongoing life time of a system.

Device

  • Model / Serial
    Model Catalog: 5904656 (Lot serial: 28022); Model Catalog: 7728392 (Lot serial: > 10 numbers contact mfg.); Model Catalog: 5904656 (Lot serial: 28069); Model Catalog: 7007755 (Lot serial: 32070); Model Catalog: 10848283 (Lot serial: NA); Model Catalog: 10848282 (Lot serial: NA); Model Catalog: 10848281 (Lot serial: NA); Model Catalog: 10848280 (Lot serial: NA); Model Catalog: 10848355 (Lot serial: NA); Model Catalog: 10848353 (Lot serial: NA); Model Catalog: 5904441 (Lot serial: NA); Model Catalog: 10280959 (Lot serial: NA); Model Catalog: 5904466 (Lot serial: NA); Model Catalog: 7555357 (Lot serial: 53097); Model Catalog: 7555357 (Lot serial: 53081); Model Catalog: 7555357 (Lot serial: 53088); Model Catalog: 7555357 (Lot serial: 53101); Model Catalog: 5904656 (Lot serial: 28048); Model Catalog: 10094141 (Lot serial: > 10 numbers contact mfg.); Model Catalog: 10094137 (Lot serial: > 10 numbers contact mfg.); Model Catalog: 7727717 (Lot serial: 40033); Model Catalog: 7412807 (Lot serial: 35177); Model Catalog
  • Product Description
    AXIOM ARTIS BA;ARTIS ZEE BIPLANE;AXIOM ARTIS TA;ARTIS Q;ARTIS Q.ZEN;AXIOM ARTIS FA ALPHA;ARTIS ZEEGO MULTI-AXIS;AXIOM ARTIS MP DIAGNOSTIC;AXIOM ARTIS DBA;ARTIS ZEE CEILING;AXIOM ARTIS DFC MAGNETIC NAVIGATION;AXIOM ARTIS DFC;AXIOM ARTIS FC;AXIOM ARTIS DBC
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC