Recall of AUTOMATICQC CARTRIDGE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18958
  • Event Risk Class
    III
  • Event Initiated Date
    2016-09-15
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has confirmed that the lots of aqc listed have been incorrectly value assigned resulting in a +0.8 g/dl bias for level 1 thb on rapidpoint 400/405/500 systems. some users are reporting higher than expected results for this aqc level. aqc thb levels 2 and 3 and all other aqc parameters are not affected.

Device

  • Model / Serial
    Model Catalog: 10310323 (Lot serial: AQC/21516); Model Catalog: 10310323 (Lot serial: AQC/21416); Model Catalog: 10310323 (Lot serial: AQC/21016); Model Catalog: 10310323 (Lot serial: AQC/19516); Model Catalog: 10310323 (Lot serial: AQC/20816); Model Catalog: 10310323 (Lot serial: AQC/19616); Model Catalog: 10310323 (Lot serial: AQC/20916)
  • Product Description
    AUTOMATICQC CARTRIDGE
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC