Recall of ASP AUTOMATIC ENDOSCOPE REPROCESSOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by JOHNSON & JOHNSON MEDICAL PRODUCTS A DIV. OF JOHNSON & JOHNSON INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    76174
  • Event Risk Class
    II
  • Event Initiated Date
    2010-04-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Three customer reports have been received of elevated temperature of the high-level disinfectant solution used in the asp aer as a result of a possible temperature monitoring malfunction. if high-level disinfectants are exposed to elevated temperatures fumes may occur and result in reported incidents of respiratory reactions dizziness burning eyes and headaches by the healthcare worker.

Device

Manufacturer