Recall of ARTISTE MV - LINEAR ACCELERATOR - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    142482
  • Event Risk Class
    III
  • Event Initiated Date
    2014-06-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens became aware of an error in the pressure/temperature compensation for dose monitor 2 for electron treatments. only digital linear accelerators (linac) running control console software version 12 or 13 are affected. this error may result in a higher deviation of the dose rate in electron mode than specified. the dose monitor 1 controls the overall dose monitoring system and remains unaffected. the actual dose rate and the accumulated monitor units and thus the absolute dose are correctly displayed at the linac console.

Device

  • Model / Serial
    Model Catalog: 8139789 (Lot serial: 5414)
  • Product Description
    ARTISTE MV SYSTEM - DIGITAL LINEAR ACCELERATOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC