Recall of ARTIS ZEEGO MULTI-AXIS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    23496
  • Event Risk Class
    III
  • Event Initiated Date
    2014-06-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The hand switch to release a x-ray image at the patient table is equipped with a cable extension inside the table. a problem may occur if the connection between cable connection and hand switch is contaminated with an extraordinary large amount of fluid. in a worst case situation uncontrolled release of a x-ray image may happen. if this situation occurs the system terminates uncontrolled x-ray after a short period.

Device

  • Model / Serial
    Model Catalog: 10280959 (Lot serial: 160423); Model Catalog: 7728392 (Lot serial: 44274); Model Catalog: 10280959 (Lot serial: 160109); Model Catalog: 10094135 (Lot serial: 136104); Model Catalog: 10094135 (Lot serial: 136099); Model Catalog: 10094135 (Lot serial: 136226); Model Catalog: 10094135 (Lot serial: 135367); Model Catalog: 7008605 (Lot serial: 55433); Model Catalog: 7008605 (Lot serial: 55402); Model Catalog: 10094139 (Lot serial: > 10 numbers contact mfg); Model Catalog: 7555365 (Lot serial: > 10 numbers contact mfg); Model Catalog: 10094137 (Lot serial: > 10 numbers contact mfg); Model Catalog: 10094141 (Lot serial: 153372); Model Catalog: 7728392 (Lot serial: 153372); Model Catalog: 10094141 (Lot serial: 153369); Model Catalog: 7728392 (Lot serial: 153369); Model Catalog: 10094141 (Lot serial: 153810); Model Catalog: 7728392 (Lot serial: 153810); Model Catalog: 10094141 (Lot serial: 153854); Model Catalog: 7728392 (Lot serial: 153854); Model Catalog: 10094141 (Lot serial: 153275); Model Catalog:
  • Product Description
    ARTIS ZEEGO
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC