Recall of ARTHROSCOPIC IRRIGATION SET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65076
  • Event Risk Class
    II
  • Event Initiated Date
    2013-05-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    An internal investigation led to the discovery of pinhole defects in the clear bottom film of the packaging of the two lead arthroscopic irrigation set (code 2c4030) for use with arthromatic plastic containers. these pinholes are not likely to be detected by the end user prior to use. pinholes could compromise the sterile barrier properties of the packaging.

Device

  • Model / Serial
    Model Catalog: 2C4030 (Lot serial: FROM R06K01067 TO R12F11067)
  • Product Description
    TWO LEAD ARTHROSCOPIC IRRIGATION SET
  • Manufacturer

Manufacturer