Recall of ARGUS EPIC CAMERA

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    53441
  • Event Risk Class
    II
  • Event Initiated Date
    2013-10-31
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A recent investigation has determined that visual inspections of the fork joints behind the detector that is conducted during six month preventive maintenance (pm) may not detect small cracks or deficiencies in the fork weldment.

Device

  • Model / Serial
    Model Catalog: 2145-3007A (Lot serial: N/A)
  • Product Description
    ARGUS EPIC CAMERA
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC