Recall of ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM - MODEL 5000 CONTROL MODULE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BARD CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67747
  • Event Risk Class
    III
  • Event Initiated Date
    2015-11-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Bard has identified arctic sun 5000 temperature management systems with specific serial numbers that may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system clock and static random access memory (sram). this premature coin cell battery drainage could render your device unresponsive upon system startup.

Device

  • Model / Serial
    Model Catalog: 5000-00-00 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    ARCTIC SUN 5000 TEMP MANAGEMENT SYSTEM
  • Manufacturer

Manufacturer