Recall of AQURE SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by RADIOMETER MEDICAL APS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    68808
  • Event Risk Class
    II
  • Event Initiated Date
    2015-12-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Radiometer has been made aware that the aqure system may display the sample type as "blank" if the sample type is not specified by the user. as multiple result values for the same patient and the same parameter can be displayed side by side on the screen there is a potential risk of misinterpretation of results the user may confuse the blank/unspecified sample type with the content of a specified sample type field.

Device

  • Model / Serial
    Model Catalog: 933-599 (Lot serial: N/A)
  • Product Description
    AQURE SYSTEM