Recall of APLIO XG ULTRASOUND SYSTEM - MAIN

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by TOSHIBA CANADA MEDICAL SYSTEMS LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21527
  • Event Risk Class
    II
  • Event Initiated Date
    2010-06-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The aplio xg ultrasound system (ssa-790a) does not conform to one requirement of clause 15b power supply section of iec standard 60601-1 2nd edition.

Device

  • Model / Serial
    10 NUMBERS CONTACT MFG")" v-b-tooltip> Model Catalog: SSA-790A (Lot serial: ">10 NUMBERS CONTACT MFG")
  • Product Description
    APLIO XG ULTRASOUND SYSTEM - MAIN
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC