Recall of ANACONDA STENT GRAFT SYSTEM - FLARED LEG WITH DELIVERY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIVANOVA CANADA CORP..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    45093
  • Event Risk Class
    II
  • Event Initiated Date
    2017-11-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Vascutek has recived two complaints whereby a small section of the delivery system sheath became detached from the collar during deployment of the anaconda longer leg iliac stent device from the delivery system. should this occur users should refer to the bailout procedure in the instructions for use (ifu) to complete the deployment.

Device

  • Model / Serial
    Model Catalog: FL1223X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1221X170 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1221X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1219X170 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1219X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1217X170 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1217X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1215X170 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1215X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1213X170 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: FL1213X150 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: TL1210X170 (Lot serial: 2002380298); Model Catalog: TL1210X150 (Lot serial: 2002380298); Model Catalog: TL1210X170 (Lot serial: 2002233717); Model Catalog: TL1210X150 (Lot serial: 2002233717); Model Catalog: TL1210X170 (Lot serial: 2002224604); Model Catalog: TL1210X150 (Lot serial: 200222460
  • Product Description
    ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (FLARED);ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (TAPERED);ANACONDA LONGER LEG ILIAC STENT GRAFT SYSTEM (STRAIGHT)