Recall of ANACONDA STENT GRAFT SYSTEM - BIFURCATE BODY STENT WITH DELIVERY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIVANOVA CANADA CORP..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    23740
  • Event Risk Class
    II
  • Event Initiated Date
    2013-11-01
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Vascutek has received 3 complaints (with an occurrence rate of 0.1%) where the release wire of an anaconda body delivery system has fractured following stent deployment. two of these three incidents resulted in conversion to open repair. the manufacturer is recalling all anaconda bodies. previously implanted stents are not affected by this recall.

Device

  • Model / Serial
    Model Catalog: B34 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B32 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B30 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B19 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B25 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B23 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B21 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: B28 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: OLB34 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB32 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB30 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB21 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB25 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB23 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: OLB28 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    ANACONDA BIFURCATE BODY STENT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC