Recall of ALLURA XPER FD20 SYSTEM - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    47714
  • Event Risk Class
    III
  • Event Initiated Date
    2014-12-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that a mechanical noise/tick might occur when moving the xy-carrier over the x-rail coupling. this problem can occur when an "old" version of coupling block with m5 bolts is installed in the system.

Device

  • Model / Serial
    Model Catalog: ALLURA XPER FD20 (Lot serial: ALL)
  • Product Description
    ALLURA XPER FD20 SYSTEM - MAIN UNIT