Recall of ALLURA XPER FD20 OR TABLE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    49935
  • Event Risk Class
    III
  • Event Initiated Date
    2016-03-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips healthcare has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with health canada safety code 20a and some provincial regulations. in certain circumstances a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound as required in health canada safety code 20a and some provincial regulations. no injuries attributed to the problem are reported.

Device

  • Model / Serial
    Model Catalog: 000454 (Lot serial: 33); Model Catalog: 000454 (Lot serial: 146); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 607); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 171); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 97); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 154); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 215); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 40); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 489); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 405); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 449); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 306); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 307); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 168); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 567); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 283); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 57); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 52); Model Catalog: ALLURA XPER FD20/10 (Lot serial: 37); Model Catalog: ALLURA XPER FD
  • Product Description
    ALLURA XPER FD20 OR TABLE