Recall of ALLURA XPER FD10/10 - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    49936
  • Event Risk Class
    II
  • Event Initiated Date
    2014-09-16
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that if a footswitch is frequently used on an anti-fatigue mat on a not flat surface or in the pedestal the footswitch pedals may get bent. this might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

Device

  • Model / Serial
    Model Catalog: ALLURA XPER FD10/10 (Lot serial: 215); Model Catalog: 0708MD02 (Lot serial: 194); Model Catalog: 0708MD02 (Lot serial: 126); Model Catalog: 0708MD02 (Lot serial: 107); Model Catalog: 0708MD02 (Lot serial: 78); Model Catalog: 0708MD02 (Lot serial: 69); Model Catalog: 0708MD02 (Lot serial: 65); Model Catalog: 0708MD02 (Lot serial: 13); Model Catalog: ALLURA XPER FD20 (Lot serial: >10 numbers contact mfg.); Model Catalog: 0708MD02 (Lot serial: 215); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 527); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 567); Model Catalog: ALLURA XPER FD20/20 (Lot serial: 607); Model Catalog: ALLURA XPER FD10 (Lot serial: >10 numbers contact mfg.)
  • Product Description
    ALLURA XPER FD10/10 - MAIN UNIT