Recall of ALDOLASE MANUAL METHOD

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by RANDOX LABORATORIES LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    25347
  • Event Risk Class
    III
  • Event Initiated Date
    2016-05-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Customers using a randox rx imola instrument only will not be able use randox aldolase reagent (ad189) with the instrument specific application provided. the parameters shown on the screen shots on the instrument specific application sheet for aldolase assay on the rx imola instrument are incorrect. the settings do not allow for absorbance change therefore calibration is not possible and the assay cannot be run as a test. affected customers will receive an updated instrument specific application sheet for aldolase with the correct parameters.

Device

  • Model / Serial
    Model Catalog: AD189 (Lot serial: ALL)
  • Product Description
    ALDOLASE MANUAL METHOD
  • Manufacturer

Manufacturer