Recall of AIXPLORER ULTRASOUND SYSTEM - CURVED ARRAY PROBE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CHRISTIE INNOMED INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27049
  • Event Risk Class
    II
  • Event Initiated Date
    2012-02-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Supersonic imagine has identified an error in the user's guide regarding the list of recommended cleaners/disinfectants. the disinfectants recommended for the se12-3 robe are not all high level. disinfectants. the se12-3 is used in semi-critical application so it should be cleaned and high level disinfected after use.

Device

  • Model / Serial
    Model Catalog: SSIP90030 (Lot serial: > 10 serial # contact mfg); Model Catalog: SSIP90068 (Lot serial: > 10 serial # contact mfg); Model Catalog: SSIP90019 (Lot serial: >10 serial # Contact Mfg); Model Catalog: SSIP09002 (Lot serial: >10 serial # contact mfg)
  • Product Description
    Supersonic Aixplorer Ultrasound System
  • Manufacturer