Recall of AGILITY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ELEKTA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    138514
  • Event Risk Class
    II
  • Event Initiated Date
    2013-12-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During the treatment table gantry and collimator calibration procedures it is possible that the user type incorrect values or invert a value. if quality assurance (qa) procedures are not conducted after the calibration it is possible that the error will go unnoticed.

Device

  • Model / Serial
    Model Catalog: MRT 16701 (Lot serial: N/A); Model Catalog: 1010318 (Lot serial: N/A); Model Catalog: MRT 6011 (Lot serial: N/A); Model Catalog: MRT 6001 (Lot serial: N/A); Model Catalog: 571337 (Lot serial: N/A)
  • Product Description
    Integrity Software 3.0.0