Recall of AERONOX NO DELIVERY AND ANALYSIS SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PULMONOX MEDICAL INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32257
  • Event Risk Class
    II
  • Event Initiated Date
    2003-06-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Failure of the no2 sensor used in the aeronox. this is due to the open loop feedback condition caused by small input offset voltages of two operational amplifiers connected when the unit powers off. dk.

Device

  • Model / Serial
    Model Catalog: (Lot serial: )
  • Product Description
    AERONOX NO DELIVERY AND ANALYSIS SYST
  • Manufacturer

Manufacturer

  • Manufacturer Address
    TOFIELD
  • Source
    HC