Recall of ADVIA CENTAUR SYSTEM - PROSTATE-SPECIFIC ANTIGEN (PSA) ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16231
  • Event Risk Class
    III
  • Event Initiated Date
    2015-12-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Not meeting the current high-dose hook effect expectation in the instructions for use (ifu). the advia centaur systems psa assay range is 0.01-100 ng/ml (ug/l). the advia centaur systems psa ifu states that patient samples with high total psa levels can cause a paradoxical decrease in the relative light units (rlus) (high-dose hook effect). in this assay patient samples with total psa levels as high as 10000 ng/ml (ug/l) will assay greater than 100 ng/ml (ug/l).

Device

  • Model / Serial
    Model Catalog: 118157 (Lot serial: ending in 263 and above); Model Catalog: 118156 (Lot serial: ending in 263 and above)
  • Product Description
    ADVIA CENTAUR SYSTEM - PROSTATE-SPECIFIC ANTIGEN
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC