Recall of ADVIA CENTAUR CP SYSTEM - ENHANCED ESTRADIOL (EE2) ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    112708
  • Event Risk Class
    I
  • Event Initiated Date
    2016-01-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics is conducting a correction for all lots of advia centaur enhanced estradiol dimension vista loci estradiol immulite/immulite 1000 estradiol and immulite 2000 estradiol. siemens has confirmed the drug fulvestrant (faslodex) may cause falsely elevated estradiol results in the assays listed.