Recall of ADVANCED PERFUSION SYSTEM 1 - BASE UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by RYAN MEDICAL DISTRIBUTORS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    38812
  • Event Risk Class
    II
  • Event Initiated Date
    2012-11-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Terumo cardiovascular systems has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. occlusion malfunctions include stiff/difficult to adjust occlusion loss of occlusion setting unequal roller-to-roller occlusion and inability to adjust the occlusion.

Device

  • Model / Serial
    Model Catalog: 801763 (Lot serial: >10 numbers contact mfg)
  • Product Description
    Terumo Advanced Perfusion System 1 -Base Unit
  • Manufacturer

Manufacturer