Recall of ACT 5DIFF SERIES HEMATOLOGY ANALYZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27829
  • Event Risk Class
    II
  • Event Initiated Date
    2011-02-24
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    There is a potential for erroneously high platelet (plt) rbc mpv and hematocrit (hct) results to be generated. the erroneous results may occur on the first aspiration of a tube and may occur without instrument generated messages. testing has demonstrated an average probability of occurrence of erroneous elevated platelet counts on the first aspiration of a specimen tube of 0.6% with a 95% confidence interval of 0.21% to 0.99%.

Device

  • Model / Serial
    Model Catalog: 175356 (Lot serial: ALL)
  • Product Description
    ACT 5DIFF SERIES HEMATOLOGY ANALYZER
  • Manufacturer

Manufacturer