Recall of ACL 8000 COAGULATION ANALYZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28705
  • Event Risk Class
    II
  • Event Initiated Date
    2010-11-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Beckman coulter has confirmed sporadically lowered aptt results outside of the clia allowable error limit of 15% when using these il locked test groups.

Device

  • Model / Serial
    Model Catalog: 0000880011 (Lot serial: 03060101 to 09100677); Model Catalog: 0000990011 (Lot serial: 00040105 to 09031882); Model Catalog: 0000880311 (Lot serial: 05010100 to 10111176); Model Catalog: 0001000311 (Lot serial: 05010100 to 10111779); Model Catalog: 01000011 (Lot serial: 02100101 to 06020501)
  • Product Description
    ACL 8000 COAGULATION ANALYZER
  • Manufacturer

Manufacturer