Recall of ACCUTREND GC MONITOR (GLUCOSE/CHOLESTEROL)

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ROCHE DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    23578
  • Event Risk Class
    II
  • Event Initiated Date
    2007-12-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Since may 2005 meters were sold with old version of operator manual instead of updated version. risk of infection and potential for erroneous results when instructions for multiple-patient use are not followed.

Device

  • Model / Serial
    Model Catalog: 1418246 (Lot serial: ALL LOTS)
  • Product Description
    ACCUTREND GC (MULTIPLE PATIENT USE)
  • Manufacturer

Manufacturer

  • Manufacturer Address
    LAVAL
  • Manufacturer Parent Company (2017)
  • Source
    HC