Recall of ACCESS SYSTEM - TOTAL T4 ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    24181
  • Event Risk Class
    III
  • Event Initiated Date
    2012-04-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During an internal study some patient samples with total t4 concentrations above the measuring range of the access total t4 assay when diluted according to the access total t4 product instructions for use using one volume of sample with an equal volume of access total t4 calibrator s0 (zero) were falsely elevated up to 22%.

Device

  • Model / Serial
    Model Catalog: 33800 (Lot serial: >10 LOT NUMBERS CONTACT MFG)
  • Product Description
    ACCESS TOTAL T4 REAGENT (ACCESS SYSTEM)