Recall of ACA STAR DISCRETE CLINICAL ANALYZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DUPONT CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    129912
  • Event Risk Class
    III
  • Event Initiated Date
    1998-10-06
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Bar code on sal-c carton contains an incorrect method mneumonic as the method name comes up up instead of sal the instrument will not calibrate automatically.

Device

  • Model / Serial
    Model Catalog: (Lot serial: 7KD084 8BD057)
  • Product Description
    ACA STAR DISCRETE CLINICAL ANALYZER
  • Manufacturer

Manufacturer

  • Manufacturer Address
    ONTARIO
  • Source
    HC