Safety Alert for Trade Name: Vitros * ECI Technical Name: Tampons, Diluents and other solutions for laboratory analysis ANVISA Registration Number: 10132590378 Hazard Class: I Affected Model: 2 vials x 28 ml Serial Numbers Affected: SKU: 1072693 | Lot: 4492 or higher

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Johnson & Johnson do Brasil Ind. e Com. de Prod. para Saúde Ltda; Ortho-Clinical Diagnostics..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2377
  • Date
    2017-09-06
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    If your VITROS 3600 or 5600 system can not detect the presence of VITROS signal reagent packets, contact the Ortho Care ™ Technical Solutions Center for immediate assistance. An Ortho trained service representative will contact you to schedule the MOD C3 installation. • Please send this notification if affected batches have been distributed to a different location. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Complementary Information: - Date of identification of the problem by the company: 09/08/2017 - Date of notification notice to Anvisa: 09/09/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Reason
    Sensors in the vitros system may not detect the base of the updated vitros signal reagent packages due to a larger indentation in the base of the reagent package. as a result, some vitros 3600 or 5600 systems may not be able to continuously detect vitros signal reagent packs in the load positions or during the test in the position in use when using lot 4492 or higher.
  • Action
    Field Action Code TC2017-157 triggered under the responsibility of Johnson & Johnson do Brasil Ind. And Com. De Prod. for Health Ltda. Perform Field Correction.