Safety Alert for Trade name: SOMATOM Computerized Tomography System. Technical name: CT Scanner. . Model: SOMATOM Emotion Registration No .: 10345161972, Risk Class: III; Model: SOMATOM Spirit; Risk Class: III; Registration No: 10345162002 Hazard Class: III. Model: SOMATOM Scope Power, Registration No: 10345161975, Risk Class III :. Model: SOMATOM Perspective, Registration Number: 10345162001, Hazard Class: III

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Siemens Healthcare Diagnósticos Ltda; Siemens Healthcare GmbH.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2506
  • Date
    2018-03-02
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    The problem described will not occur when using a lateral topogram instead of a topograph or pa. Consequently, we recommend the use of topograms in lateral position for all head acquisitions. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link
  • Reason
    Siemens has identified possible incorrect amphole current calculations by the care dose4d algorithm when used on head tomography based on topograms pa (posterior-anterior) or ap.
  • Action
    Field Action Code CT076 / 17 / S & CT077 / 17 / S triggered under the responsibility of Siemens Healthcare Diagnósticos Ltda. The company will send a letter to the customer.