Safety Alert for Trade name: Linear Digital Primus Accelerator and Linear Accelerator Oncor. Technical name: Linear accelerator. ANVISA registration number: 10234230073; 10234230125. Class of risk: III. Affected Model:. Serial number affected: See distribution map

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Siemens Healthcare Diagnósticos Ltda.; Siemens AG; Siemens Medical Solutions USA, Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2521
  • Date
    2018-03-16
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    During a manual movement, the user has to monitor the system all the time and if the gantry is approaching the patient too much, stop the movement by releasing the safety switch pressed (if operated by manual control) or the keys pressed (if operated Control Console keyboard). If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 02/14/2018 - Date of notification notice to Anvisa: 03/15/2018 The company that owns the affected product is responsible for contacting its customers in due time. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Reason
    Siemens informs you that the control console software will be updated to version 9.2.400. this update aims to reduce the likelihood of a collision by automatically applying treatments sequenced with automatic movements of the gantry and / or treatment table using the simtec ™ automatic field matching option, so siemens updated the software solution "protection automatic field sequence motion (afs) ". the previous "afs motion protection" software only considered the critical adjustments of isocentric and eccentric table positions, but not a lateral displacement of the treatment table.
  • Action
    Field Action Code TH009 / 17 / S & TH010 / 17 / S triggered under the responsibility of Siemens Healthcare Diagnósticos Ltda. Will make field correction

Manufacturer