Safety Alert for Trade name: Linear Accelerator Elekta. Technical Name: Linear Accelerator. ANVISA registration number: 80569320007. Risk class: III. Affected Model: Integrity ™. Serial numbers affected :. INTEGRITY SOFTWARE / 16966-001 / 3.2.0 ;. INTEGRITY SOFTWARE / 30000301-001 / 3.2.0. INTEGRITY SOFTWARE / 20391-001 / 3.2.0. INTEGRITY SOFTWARE / 30000515-002 / 3.2.0. INTEGRITY SOFTWARE / 30000683-001 / 3.2.0

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Elekta Medical Systems Comércio e Serviços Radioterapia Ltda; Elekta Ltd.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2473
  • Date
    2018-01-18
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    Read and follow Important Safety Notice 200-01-502-053. Send your signed Elekta representative as soon as possible the signed document receipt. Allow Elekta Service representatives to access to install the modification when it is launched. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 10/20/2017 - Date of notification notice to Anvisa: 01/18/2018 The company that owns the affected product is responsible for contacting its customers in due time. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Reason
    Potential for positional errors after automatic table movement (atm) elekta has identified that it is possible to obtain a positional error with the precise treatment table ™ after automatic table movement. this may occur if there is an undetected failure of the position sensors. this problem was previously reported through 200-01-204-011 notification. the precise treatment table is equipped with sensors that achieve a high degree of reliability when reading the table position. if there is an undetected fault on one of these sensors, a positional error can be obtained with the precise treatment table ™ after automatic table movement when using the xvi and mosaiq ™. the system has a software check to detect large positional errors resulting from sensor malfunction. in these cases, it is possible to position the treatment table with errors greater than 5 mm without any inhibitions. this may occur if automatic table movement is used. if automatic table movement is performed by iguide, positional error is detected.
  • Action
    Field Action Code FCA-EL-0004 triggered under the responsibility of Elekta Medical Systems Comércio e Serviços Radiotherapy Ltda. Will make correction in the field.