Safety Alert for HOMECHOICE FOR AUTOMATED PERITONEAL DIALISE. HOMECHOICE SYSTEM - VERSION 8.51

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Baxter Hospitalar Ltda.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    679
  • Date
    2002-10-03
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    BRAZIL - 29.10.2002 - The automated Peritoneal Dialysis System for Homechoice and Homechoice PRO is registered in Brazil under the number 10068390123. According to company information (letter dated September 27, 2002), there are no reports of physical damage associated with this situation. Baxter is sending important information to be observed with the Homechoice System, software version 8.51, launched in June 2002 to notify the clinical and patient staff of home-based DPA-directed actions to reduce the likelihood of a possible excessive infusion of volume , which could occur under certain conditions.
  • Reason
    "it was found that stopping or advancing a manual drainage during an infusion phase could result in excessive infusion of volume. therefore, if the patient requires manual drainage during the filling phase, it is important that drainage is performed without interruption, with purpose of preventing excessive volume infusion. these instructions apply only to versions 8.51 and 8.52 of the homechoice system software.
  • Action
    Routing of safety information to customers about the Homechoice System. Proposal for corrective actions (BAXTER): - Send letters with information to affected patients and to coordinators DE Home Care; -Send revised instructions from the patient manual to Homechoice users, help reports, and an operator's manual. -Correct the possibility of excessive infusion of volume with the next version of the software. For further information please contact us on 0800 12 55 22, option 1.