Safety Alert for GSP Genetic Screening Processor Technical Name: ANVISA Registration Number: 10298910108 Hazard Class: I Affected Model: 2021-0010 Affected serial numbers: 20210154, 20210207, 20210171, 20210163, 20210161, 20210208.

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by PerkinElmer do Brasil Ltda.; Wallac Ou.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2193
  • Date
    2016-09-02
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    Defective insulation will be replaced during the next scheduled periodic maintenance visit or service visit. The visit can also be scheduled if requested by the client.
  • Reason
    We are aware that the insulation of the heated incubator door is defective in the instrument serial numbers listed above due to a change in the insulation production method. the film may come loose from the outer surface of the insulation, causing a mechanical motion error or a temperature control error. due to the defect, the instrument's operations will stop and it will require repair by perkinelmer field service engineers. instrument performance is only affected when error occurs in mechanical movement or temperature control. the operation of the heated incubator is involved during the incubation steps of the gsp neonatal galt (3303-0010 and 3303-001u) and gsp neonatal biotinidase (3307-0010, 3307-001u) assays which require a high temperature (+ 37 ° c ). the functionality of the heated incubator door is also checked during the instrument start procedure.
  • Action
    Field Action Code 2016.06.005461 triggered under the responsibility of the company PerkinElmer do Brasil Ltda. Company will make correction in the field.