Safety Alert for Digital Mammography Unit, MAMMOMAT Inspiration - Model: MAMMOMAT Inspiration - Record: 10234230164 - Risk Class III - High Risk - Technical Name: Mammography Unit - Series: 3064; 5047

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Siemens Healthcare Diagnósticos Ltda.; Siemens AG.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    1462
  • Date
    2014-10-20
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    The company recommends "(...) inspect the safety switch before each biopsy until the problem is solved and then carefully follow the instructions in the user manual - safety switch chapter. The customer will be advised to wait for the Service Team to schedule field corrections and to file the letter with the Operator's Manual. In addition, Siemens advises to inform all personnel of the organization who need to be aware of this problem and act accordingly. Send this safety information to any other institutions that may be affected by this measure. "
  • Reason
    Problem: according to the company: "stereotactic biopsy devices for mammomat inspiration mammography systems may have integrated a safety switch that causes a functionality failure. the pin included in the safety switch may not exert sufficient pressure on the safety circuit to prevent movement. the needle positioner can move even with the safety switch on the side. if this occurs with the needle already inserted, it can lead to an unintentional injury to the patient. ".
  • Action
    Company will perform Correction in the field. Code: XP053 / 14 / S

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    ANVSANVISA