Safety Alert for Commercial Name: Intellivue Patient Monitor. Technical Name: Vital Signs Monitor. ANVISA registration number: 10216710215. Risk class: III. Affected model: MX400, MX450, MX500, MX550, MX700, MX800. Serial numbers affected: DE5630GA7B; DE5630GHKT; DE5630GHKV; DE5630GHKW; DE5630GHLB; DE5630GHLC; DE5630GHLE; DE5630GHLF; DE5630GHLG; DE5630GHLH; DE5630GHLJ; DE5630GHLK; DE5630GHLL; DE5630GHLM; DE5630GHLN; DE5630GHLP; DE5630GHLT; DE5630GHLU; DE5630GHLY; DE5630GHM4; DE5630GHM5; DE5630GHM7; DE5630GHM9; DE5630GHME; DE5630GHMV; DE5630GHMW; DE5630GHMY; DE5630GHMZ; DE5630GHN2; DE5630GHN3; DE5630GEC3; DE5630GDZE; DE5630GEB9; DE5630GEBM; DE5630GEB4; DE5630GEBN; DE5630GEBK; DE5630GDN6; DE5630GEC2; DE5630GEBH; DE5630GEC0; DE5630GDP5; DE5630GEC1; DE5630GEBC

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Philips Medical Systems Ltda.; Philips Medizin Systeme Böblingen GmbH.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2381
  • Date
    2017-09-21
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    Until the user software is updated, you must make sure that the INOP "ECG Check Cable" or "ECG Noisy Elec xx" is handled using the corresponding IFU instructions . Review this information with all team members who are using the conventional 12-lead ECG from the IntelliVue Patient Monitors and need to be aware of the contents of this release. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 08/21/2017 - Date of notification notice to Anvisa: 09/21/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Reason
    The st segments of a 12-lead ecg may be distorted and the st value may be inaccurate when an affected monitor automatically activates ecg filtering in the event of an inop condition "ecg check cable" or "ecg noisy elec xx" [ecg noise electrode with noise] is triggered and displayed. these inops are triggered when the monitor detects a low impedance between the ecg lead wires and the cable insulation in the lead set that may have been caused by mechanical damage or fluid ingress.
  • Action
    Field Action Code FCO86201799 triggered under the responsibility of the company Philips Medical Systems Ltda. Will perform field correction.

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    ANVSANVISA