Safety Alert for Commercial name: Angiography equipment Artis. Technical Name: Angiography Equipment. ANVISA registration number: 10345162023. Risk class: III. Model affected: Artis Zee ceiling. Serial numbers affected: 17582

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Siemens Healthcare Diagnósticos S.A; Siemens Healthcare GmbH.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    2396
  • Date
    2017-10-19
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    There is no need for retesting in any patient. This is a possible hardware defect that does not influence the treatment of patients. In specific cases and if necessary, you can cancel or restart the clinical treatment of a patient or transfer it to a system in working condition. It is recommended that the client notify and immediately instruct all teams that need to be aware of this issue. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 24/082017 - Date of notification notice to Anvisa: 10/19/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner in order to ensure the effectiveness of the ongoing Field Action. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Reason
    The wireless radar trigger pedal may fail due to impacts from external factors such as electrostatic discharge exceeding a given intensity. problem-solving actions (redefining the affected set) have been implemented, but they have no effect in this case.
  • Action
    Field action AX009 / 17 / S & AX013 / 17 / S Code triggered under the responsibility of the company Siemens Healthcare Diagnósticos SA Will make field correction.