Safety Alert for ADVIA Centaur HCY (HOMOCYSTEIN) ANVISA Registry No. 10345160332.

According to Agência Nacional de Vigilância Sanitária (ANVISA), this safety alert involved a device in Brazil that was produced by Siemens Healthcare Diagnostics Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    1276
  • Date
    2013-07-29
  • Event Country
  • Event Source
    ANVISA
  • Event Source URL
  • Notes / Alerts
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Extra notes in the data
    Upon receipt of the communication, the customer should discontinue the use of 1:10 dilutions of samples with the ADVIA Centaur Homocysteine ​​(HCY) assay run on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP systems. Ensure that 1:10 dilution will not be selected manually or automatically when performing a dilution. See the sections on Setting Automatic Dilution and Introduction to Dilution Options in the ADVIA Centaur System Operator's Guide. Customers can continue to use the 1: 2 dilution onboard or manual for samples with results outside the assay range. Siemens has confirmed that performance for this level of dilution meets the recovery described in the Instructions for Use.
  • Reason
    Percent recovery for 1:10 dilutions of diluted patient samples is below that indicated in the instructions for use.
  • Action
    Notice to the Client. Access to alert message: http: //portal.anvisa.gov.br/wps/wcm/connect/77db1c8040881b8eb903bd47eaaaed7e/Carta_aos_Clientes.pdf? MOD = AJPERES