Recall of Ysio, Luminos Agile, Luminos dRF and Uroskop Omnia Systems with software Versions VB10 and VC10 (diagnostic digital x-ray system)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01037-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-10-09
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    During a rad (radiography) examination with automatic exposure control, a highly unlikely malfunction can result in radiation continuing longer than required. this malfunction occurs when the generator stops the tube power too late and therefore too much dose is applied during an acquisition. this may also result in an overexposed image that is not of diagnostic quality, so that the examination needs to be repeated. statistically this malfunction only occurs once every 7000 images and only if images are taken in automatic exposure mode.
  • Action
    It is not possible for the user to prevent this malfunction. The exposure time with the use of automatic exposure control is limited to 500ms, which is the maximum time that unwanted radiation will continue. The exposure can be interrupted by the user at anytime by releasing the exposure button. Siemens has implemented a field modification that will resolve the issues. The Australian roll out of this field modification is planned for completion in Q4 2013. This action has been closed-out on 11/02/2016.