Recall of Wireless Footswitch used in conjunction with AXIOM Artis, Artis Zee and Artis Q Systems

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00544-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-05-20
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    If the wireless footswitch gets disconnected from the loading station due to an unintentional pull of the plug connection, the plug can become damaged.As a consequence, this could have the implication that the wireless footswitch can no longer be charged. safe removal of the plug connection requires the interlocking mechanism to be loosened.
  • Action
    The cabling that is currently being used will be replaced for all affected systems that are installed. The new solution has a coupling device that automatically unlocks upon draw. In addition, Siemens will provide an addendum for the operating manual that describes the appropriate handling required by the user.

Device

  • Model / Serial
    Wireless Footswitch used in conjunction with AXIOM Artis, Artis Zee and Artis Q SystemsCharging stations (catalogue no. 04787763) with serial numbers <10,000 are affectedARTG Number: 145549