Recall of Volume Imaging System (XVI), software versions 4.2.1 (Electronic Imaging Device (EID) used with radiation therapy treatment)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Elekta Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01291-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2013-12-03
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Xvi can incorrectly calculate the target position of the treatment table, because of an error due to a specific sequence of events that does not occur very frequently.If the user does not monitor the patient and treatment table movements, and keeps the table automatic setup (asu) buttons pressed, the treatment table continues to move potentially resulting in a fatal collision with the patient. it only stops when a touchguard activates. this error can cause a collision between the patient and an external beam limiting device (bld) if:- the external bld does not operate with the touchguard attached, or - the external bld extends out more than the touchguard. when the error occurs, it is not possible to get the table move assistant values in tolerance. therefore, it is not possible to apply an incorrect correction to the patient position before treatment delivery.
  • Action
    Elekta is providing work around instructions for users to follow. Elekta is advising users to monitor the Actual column of the Table Move Assistant dialog box. If the Actual is blank, do not press the ASU button on the function keypad (FKP). To correct the error, users are advised to restart the XVI, place patient in initial set up position and scan the patient again and continue the usual workflow. A software update is expected to be released in April 2014 for upgrading all XVI 4.2.1 users to XVI 4.2.2 to resolve this issue. This action has been closed-out on 17/05/2017.

Device

  • Model / Serial
    Volume Imaging System (XVI), software versions 4.2.1 (Electronic Imaging Device (EID) used with radiation therapy treatment)ARTG Number: 214673