Recall of VITEK 2 System

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Biomerieux Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01342-1
  • Event Risk Class
    Class III
  • Event Initiated Date
    2013-12-19
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    There have been recent complaints that the vitek 2 meropenem (mem, mem02n) is overcalling resistance for isolates of aeromonas hydrophila. false resistance was confirmed when isolates from customer sites were tested internally by broth microdilution (bmd), the reference method used for development of mem02n.
  • Action
    Biomerieux is advising their cusotomers to implement additional supression rules for the reporting of resistant results for Aeromonas species.