Recall of Various mass spectrometers - Quattro micro, ACQUITY TQD, Quattro Premier XE, Xevo TQ; In vitro diagnostic medical devices (IVD's)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Emergo Asia Pacific Pty Ltd T/a Emergo Australia.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2012-RN-01068-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2012-10-22
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    There is a possibility that the ion block source heater may fail. in some instances, the failure could cause short term loss or variation of sensitivity that may subsequently lead to inaccurate data. the loss of or variation in sensitivity is most pronounced in assays with shorter dwell times and may not always be detected by quality control and internal standard value monitoring.
  • Action
    Customers to contact their Waters field service engineer to order a replacement part (ion block source heater).

Device

  • Model / Serial
    Various mass spectrometers - Quattro micro, ACQUITY TQD, Quattro Premier XE, Xevo TQ; In vitro diagnostic medical devices (IVD's)Quattro microPart:VB-QMicroSerial Numbers:QAA1167, QAA412, QAA742, QAA65, QAA765, QAA802 & QAA836ACQUITY TQDPart:186003875Serial Numbers:QBA470, QBB795, QBB924 & QBB928Quattro Premier XEPart:186003250Serial Numbers:VAB1000, VAB1215, VAB610, VAB844, VAB845 & WAB998Xevo TQPart:186004490Serial Numbers: VBA626 & VBA744ARTG Number: 201727
  • Manufacturer

Manufacturer