Recall of UniCel DxC Synchron Wash Concentrate II when used with Thermo-Fisher Lithium reagent (Ref A19611). An in vitro diagnostic medical device (IVD).

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Beckman Coulter Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-01619-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-12-13
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Beckman coulter has determined that certain lots of dxc wash concentrate ii can cause a failure in calibration of thermo-fisher lithium reagent (ref a19611). the consequence of calibration failure does not allow for patient results to be generated. lithium calibration failure can potentially lead to a delay of reporting lithium results. patient results reported under acceptable calibration and quality control evaluation are not affected by this issue.
  • Action
    Beckman Coulter is advising Thermo-Fisher Lithium assay users to discontinue use of affected DxC Wash Concentrate II lots and discard any unused reagent bottles. Replacement product will be provided. Reassessment of past results is not required.

Device

  • Model / Serial
    UniCel DxC Synchron Wash Concentrate II when used with Thermo-Fisher Lithium reagent (Ref A19611). An in vitro diagnostic medical device (IVD).Wash Concentrate IIReference Number: 445865Lot Numbers: Z505212 (expiry 30 April 2017), Z507201 (expiry 30 June 2016), Z511202 (expiry 31 October 2017)ARTG Number: 245295
  • Manufacturer

Manufacturer