Recall of Ultrafilter U9000 for use with AK96 and AK98 v1 dialysis machines

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Baxter Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00243-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2018-03-27
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Baxter healthcare corporation has identified potential leaks during regular clinical use of the ultrafilter u9000. the leaks are caused by cracks near the header cap due to the stress of repeated disinfection cycles over the lifetime of the product. the current instructions for use (ifu) state the ultrafilter has a maximum lifetime usage of 90 days, or 150 heat disinfection cycles, with a maximum of 8 sodium carbonate and 12 sodium hypochlorite disinfection cycles. in order to reduce the patient risk associated with the occurrence of leaks, the ifu will be updated to reduce the maximum lifetime usage to 60 days, or 100 heat disinfection cycles, with a maximum of 8 sodium carbonate and 12 sodium hypochlorite disinfection cycles when the ultrafilter is used in conjunction with ak96 and ak98 v1 dialysis machines. other machines using these filters have a leak detector sensor that will trigger an alarm.
  • Action
    Baxter is advising that operators may continue to use affected units with AK96 and AK98 v1 dialysis machines according to the new maximum lifetime usage criteria of 60 days, or 100 heat disinfection cycles, with a maximum of 8 sodium carbonate and 12 sodium hypochlorite disinfection cycles. If Ultrafilters currently in use with AK96 and AK98 v1 machines have exceeded any of the new maximum lifetime usage criteria, it is recommended the filter be changed. Baxter will reset the AK 96 & AK 98 v1 monitor counters to the reduced maximum usage during the next scheduled preventive maintenance.

Device

  • Model / Serial
    Ultrafilter U9000 for use with AK96 and AK98 v1 dialysis machinesProduct Code:112062Lot numbers: 6-1907-H-01 and higher (all unexpired units)ARTG Number: 131741(Baxter Healthcare - Filter, intravenous line)